MIL-DTL-38394E
3.4.7 Voltage and frequency. The transmitter shall withstand voltage and frequency
variations from 22 to 30 Vac and 380 to 420 hertz (Hz), or any combinations thereof.
3.4.8 Electromagnetic interference. Unless otherwise specified (see 6.2), the transmitter
shall meet the electromagnetic interference requirements of MIL-STD-461 for Air Force aircraft.
3.4.9 Damping. When the transmitter is subjected to an applied oil pressure sufficient to
produce an indication of 500 psi and the pressure is released from 500 to 100 psi within 0.1 seconds,
the corresponding time required for the transmitter indication to change from 500 to 100 psi shall be
between 3 and 7 seconds. There shall be no damage to the instrument as a result of the pressure
change.
3.4.10 Indicator operation. The transmitter shall be capable of operating either one or two
indicators with size 10 synchros without a change in output equivalent to more than 1 degree of arc.
3.4.11 Pressure bosses and electrical connector. The pressure bosses and electrical
connector shall withstand an applied force of 25 foot-pounds bending movement with no evidence
of damage to welds or soldering, distortion of base plates, and without the calibration shifting out
of tolerance.
3.4.12 Scale error. Scale error shall not exceed the tolerances specified in table I.
3.4.13 Friction and position error. Friction and position error shall not exceed the
tolerances specified in table I.
3.4.14 Effect of lead length. When the lead length (No. 20 American Wire Gauge
(AWG) stranded copper wire) between the transmitter and indicator synchro stators is changed
from 5 to 300 feet, the indication shall not change more than 1 degree of arc.
3.4.15 Case leakage. When a pressure of approximately 0 psi is applied to the exterior
of the transmitter case, the leak rate shall not exceed 0.10 micron cubic foot per hour when
10 percent helium is used in the filling medium and shall not exceed 1.0 micron cubic foot per
hour when 100 percent helium is used as the filling medium.
3.4.16 Sensors matching. The transmitter will remain accurate when pressure is applied
to the vent port.
3.5 Case. The case shall conform to figure 1 and shall provide a hermetically sealed
enclosure for the internal mechanism of the transmitter. Sealing of the case shall be
accomplished so that the seal will not be dependent upon material that would be affected by the
action of any atmosphere to which the transmitter may be subjected.
3.5.1 Filling medium. The filling medium shall be at least 98 percent free of dust particles
and shall contain not more than 0.006 milligram of water vapor per liter (dew point -65 ºC) at the
filling pressure. The filling medium shall be either a mixture of 88 to 92 percent nitrogen and the
remainder helium, or 100 percent helium. The absolute pressure of the filling medium in the case
shall be 12 to 14.7 psi.
8
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business